Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 11: (1 year) CAL VER /5500 INSTRUMENT: Fluke 12: (1 year) CAL VER /5500 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.7 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 191 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev. 1, 7/93). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55_11_2w.bmp # 55_12_2w.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F V 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 DISP Move switch to OFF position. 1.006 DISP 1.006 DISP Note: Erroneous result may occur if switch is not 1.006 DISP placed in OFF position before beginning tests. 1.007 HEAD {DC VOLTAGE TESTS}: 4V Range 1.008 JMPT 1.011 Fluke 12: (1 year) CAL VER /5500 1.009 PIC 55_11_2W 1.010 JMP 1.012 1.011 PIC 55_12_2W 1.012 DISP Move the Slide-Switch to the center position. 1.012 DISP Press RANGE to select the 4V Range (0.000). 1.013 JMPT 3.003 Fluke 11: (1 year) CAL VER /5500 1.014 ASK- Q 1.015 5500 0.000V S 2W 1.016 HEAD DC VOLTAGE TESTS: {4V Range (MAX)} 1.017 DISP Press "M" to select MIN MAX. 1.017 DISP Press "M" again to display MAX. 1.018 MEMI Enter UUT reading in volts: 1.019 MEME 1.020 MEMC 4 V 0.012U 2.001 MEME 2.002 HEAD DC VOLTAGE TESTS: {4V Range (MIN)} 2.003 DISP Press "M" to display MIN. 2.004 MEMI Enter UUT reading in volts: 2.005 MEME 2.006 MEMC 4 V 0.012U 3.001 DISP Press "M" for 2s to exit MIN MAX. 3.002 ASK+ Q 3.003 HEAD DC VOLTAGE TESTS: {4V Range (LoZ)} 3.004 DISP Move switch to the right position. 3.005 5500 4 0.000V 0.002U 2W 4.001 HEAD DC VOLTAGE TESTS: {40V Range (LoZ)} 4.002 DISP Press RANGE to select the 40V Range (0.00). 4.003 5500 40 0.02V 0.01U 2W 5.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 5.002 DISP Move switch to the center position. 5.003 DISP Press RANGE to select the 400V Range (0.0). 5.004 5500 400 -0.2V 0.1U 2W 6.001 HEAD DC VOLTAGE TESTS: {4V Range (HiZ)} 6.002 DISP Press RANGE for 2s to select autorange. 6.003 5500 4 3.500V 0.034U 2W 7.001 HEAD DC VOLTAGE TESTS: {40V Range (HiZ)} 7.002 5500 40 -35.00V 0.33U 2W 8.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 8.002 5500 400 100.0V 1.0U 2W 9.001 HEAD DC VOLTAGE TESTS: {600V Range (HiZ)} 9.002 5500 600 -600V 6U 2W 10.001 HEAD {AC VOLTAGE TESTS} 10.002 DISP Press SELECT for VAC. 10.003 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 10.004 DISP Press RANGE to select the 4V range (0.000). 10.005 5500 0.000V S 2W 10.006 MEMI Enter UUT reading in volts: 10.007 MEME 10.008 MEMC 4 V +0.003U 11.001 JMPT 13.004 Fluke 11: (1 year) CAL VER /5500 11.002 ASK- Q 11.003 5500 0.000V S 2W 11.004 HEAD AC VOLTAGE TESTS: {4V Range (MAX)} 11.005 DISP Press "M" to select MIN MAX. 11.005 DISP Press "M" again to display MAX. 11.006 MEMI Enter UUT reading in volts: 11.007 MEME 11.008 MEMC 4 V +0.040U 12.001 MEME 12.002 HEAD AC VOLTAGE TESTS: {4V Range (MIN)} 12.003 DISP Press "M" to display MIN. 12.004 MEMI Enter UUT reading in volts: 12.005 MEME # Lower limit should be 0.000 but firmware bug in early units allowed -0.004 12.006 MEMC 4 V -.004U +.040U 13.001 DISP Press "M" for 2s to exit MIN MAX. 13.002 ASK+ Q 13.003 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 13.004 5500 4 0.010V 0.003U 50H SI 2W 14.001 5500 4 0.010V 0.003U 400H SI 2W 15.001 HEAD AC VOLTAGE TESTS: {4V Range (LoZ)} 15.002 DISP Move switch to the right position. 15.003 5500 4 3.500V 0.070U 50H SI 2W 16.001 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 16.002 DISP Move switch to the center position. 16.003 5500 4 3.500V 0.070U 400H SI 2W 17.001 HEAD AC VOLTAGE TESTS: {40V Range (HiZ)} 17.002 DISP Press RANGE to select the 40V range (0.00). 17.003 5500 40 35.00V 0.70U 50H SI 2W 18.001 DISP Move switch to the right position. 18.002 HEAD AC VOLTAGE TESTS: {40V Range (LoZ)} 18.003 5500 40 35.00V 0.70U 400H SI 2W 19.001 HEAD AC VOLTAGE TESTS: {400V Range (LoZ)} 19.002 DISP Press RANGE to select the 400V range (0.0). # Warmup 2k resistor so 350V will not current limit 5500A. 19.003 5500 100.0V 80H SI S 2W 19.004 PORT [D1000] 19.005 5500 150.0V 80H SI S 2W 19.006 PORT [D1000] 19.007 5500 200.0V 80H SI S 2W 19.008 PORT [D1000] 19.009 5500 250.0V 80H SI S 2W 19.010 PORT [D1000] 19.011 5500 300.0V 80H SI S 2W 19.012 PORT [D1000] 19.013 5500 400 350.0V 7.0U 80H SI 2W 20.001 HEAD AC VOLTAGE TESTS: {600V Range (HiZ)} 20.002 DISP Move switch to the center position. 20.003 DISP Press RANGE to select the 600V Range (0). 20.004 5500 600 600V 14U 400H SI 2W 21.001 HEAD {CONTINUITY TEST} 21.002 DISP Move the Slide-Switch to the right position. 21.002 DISP Press RANGE for 2s to select autorange. 21.003 5500 2.000V 0.020U 2W 22.001 HEAD CONTINUITY TEST: {VCHEK} 22.002 5500 3.430V 0.033U 2W 23.001 5500 -0.500V 0.007U 2W 24.001 HEAD {RESISTANCE TESTS} 24.002 DISP Press SELECT for Ohms. 24.003 HEAD RESISTANCE TESTS: {400 Ohm Range} 24.004 5500 400 0.0Z +0.2U 2W 25.001 5500 400 1.0Z 0.2U 2W 26.001 5500 400 350.0Z 3.4U 2W 27.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 27.002 5500 4 1.000kZ 0.010U 2W 28.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 28.002 5500 40 35.00kZ 0.33U 2W 29.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 29.002 5500 400 100.0kZ 1.0U 2W 30.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 30.002 5500 4 2.900MZ 0.027U 2W 31.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 31.002 5500 40 35.00MZ 0.56U 2W 32.001 HEAD RESISTANCE TESTS: {VCHEK} 32.002 5500 1.960V 0.9% 0.002U 2W 33.001 HEAD {CAPACITANCE TESTS} 33.002 DISP Press uF 33.003 HEAD CAPACITANCE TESTS: {1uF Range} 33.004 5500 1 0.950uF 0.020U 2W 34.001 DISP Disconnect the 5500A from the UUT. 34.002 MATH MEM1 = 0.0 34.003 MEMI Enter UUT reading in microfarads: 34.004 MEME 34.005 MEMC 1 uF 0.001U 35.001 END